Magnesium

Magnesium

SKU: R3330000022

Product for use in the quantitative determination in vitro of the concentration of Magnesium in human serum, plasma and urine. The results of the test must always be interpreted in conjunction with the clinical picture.

  • Tests per Kit

    200

  • Test Category

    Minerals & Trace Elements

  • Target Analytes

    Magnesium

  • Sample Type

    Non-haemolysed human serum samples or plasma

  • Kit Components

    Reagent: 3 x 14 mL,
    Ready to use

  • Tech Docs

    Instructions for Use  |  Safety Data Sheet

  • Intended Use

    For professional use only.

  • Clinical Significance

    About 55% of the total magnesium present in the human body is found in the bone, while the remaining portion is within the cells, bound to proteins or molecules with a negative charge. About 55% of the magnesium present in the serum is free, 30% is bound to proteins and 15% is in the form of compounds bound to phosphate or citrate ions. It is of fundamental importance in oxidative phosphorylation, in glycolysis and in cellular replication. The reference method for the determination of magnesium in serum and urine makes use of atomic absorption spectrophotometry. For routine diagnostic use, colorimetric assay methods have been developed based on compounding reactions between colouring substances (calmagite, methylthymol blue, titanium yellow, non-sulphonated derivatives of xylidyl-blue, xylidyl-blue). The reagent makes use of xylidyl blue which is more sensitive than other stains. Like the other colouring substances, it also compounds cations other than magnesium, and in particular calcium, and it must therefore be associated with a chelating agent which eliminates this type of interference (1.2).

  • Test Principle

    Xylidyl-blue forms a soluble coloured compound with magnesium in an alkaline ambient, with a maximum absorbance between 505 and 520 nm. The intensity of the colour of the Mg-xylidyl-blue compound is directly correlated to the magnesium concentration in the sample. Calcium interference is eliminated by including a specific chelator in the reagent: diethylene-glycol bis (2-amino-ethylether) N,N-N',N'-tetracetic acid (EDTA).

  • Stability & Storage

    Store at 2 - 8°C and protect from direct light. When correctly stored, the reagents are stable up the expiry date reported on the label. A slight variation in the composition of the reagent may occur between batches, but this has no effect on the test results. After opening, the vial are stable 30 days if recapped immediately and protected from contamination, evaporation, direct light and stored at correct temperature.

  • Required but Not Supplied

    Multicalibrator (R0300000006)
    Normal control serum (R0400000006)
    Pathological control serum (R0400000106)